Measurement Accuracy | ||
Voltage DC | 110.000 mV | 0.025%+0.015% * |
1.10000 V | 0.025%+0.005% * | |
11.0000 V | 0.025%+0.005% * | |
110.000 V | 0.05%+0.005% * | |
300.00 V | 0.05%+0.005% * | |
Voltage AC | 20 to 40 Hz | 2% + 10 |
40 to 500 Hz | 0.5% + 5 | |
500 to 1 kHz | 2% + 10 | |
1 kHz to 5 kHz | 10% + 20 | |
Ranges: | 1.1000, 11.000, 110.00, 300V | |
Current DC | 30.000 mA | 0.01% + 0.015% * |
110.00 mA | 0.01% + 0.015% * | |
Resistance | 11.000 Ω | 0.05% + 50 mΩ |
110.00 Ω | 0.05% + 50 mΩ | |
1.1000 kΩ | 0.05% + 500 mΩ | |
11.000 kΩ | 0.1% + 10 Ω | |
Frequency | 1.00 to 109.99 Hz | 0.05 Hz |
110.0 to 1099.9 Hz | 0.5 Hz | |
1.100 to 10.999 kHz | 5 Hz | |
11.00 to 50.00 kHz | 50 Hz | |
Pressure | Accuracy from 0.025% of range using any of 29 pressure modules. Modules available for differential, gauge, vacuum, absolute, dual and high pressure. (for detailed specifications refer to pressure modules in options and accessories) |
|
Note | * (% of reading + % of full scale) |
Source Accuracy | ||
Voltage DC | 110.000 mV | 0.01%+0.005% * |
1.10000 V | 0.01%+0.005% * | |
15.0000 V | 0.01%+0.005% * | |
Current DC | 22.000 mA (Source) | 0.01%+ 0.015% * |
22.000 mA (Simulate) | 0.02% + 0.03% * | |
Resistance | 11.000 Ω | 0.01% + 20 mΩ |
110.00 Ω | 0.01% + 40 mΩ | |
1.1000 kΩ | 0.02% + 500 mΩ | |
11.000 kΩ | 0.03% + 5Ω | |
Frequency | 0.00 to 10.99 Hz | 0.01 Hz |
11.00 to 109.99 Hz | 0.1 Hz | |
110.0 to 1099.9 Hz | 0.1 Hz | |
1100 to 21999 Hz | 2 Hz | |
22.000 to 50.000 kHz | 5 Hz | |
Note | * (% of reading + % of full scale) |